Determination of organochlorine pesticides and polychlorinated biphenyls in post-mortem human lung by matrix solid-phase dispersion with the aid of response surface methodology and desirability function

Φόρτωση...
Μικρογραφία εικόνας

Ημερομηνία

Συγγραφείς

G. N. Rallis,
V. A. Sakkas,
V. A. Boumba,
Th. Vougiouklakis,
T. A. Albanis

Τίτλος Εφημερίδας

Περιοδικό ISSN

Τίτλος τόμου

Εκδότης

Elsevier

Περίληψη

Τύπος

Είδος δημοσίευσης σε συνέδριο

Είδος περιοδικού

peer reviewed

Είδος εκπαιδευτικού υλικού

Όνομα συνεδρίου

Όνομα περιοδικού

Journal of Chromatography A

Όνομα βιβλίου

Σειρά βιβλίου

Έκδοση βιβλίου

Συμπληρωματικός/δευτερεύων τίτλος

Περιγραφή

A matrix solid-phase dispersion (MSPD) method for the simultaneous determination of 20 organochlorine pesticides (OCPs) (aldrin, endrin, dieldrin, -BHC, -BHC, -BHC, -BHC, -chlordane, -chlordane, p,p -DDE, p,p -DDT, p,p -DDD, endosulfan I, endosulfan II, endosulfan sulfate, endrin aldehyde, heptachlor, heptachlor epoxide, endrin ketone and methoxychlor) and six polychlorinated biphenyl (PCB) congeners (PCB 28, 52, 101, 138, 153 and 180) in post-mortem human lung has been developed and validated. Response surface methodology (RSM) and desirability function were employed to optimize the extraction conditions of MSPD. Extraction was carried out using Florisil (2.0 g) as the sorbent material as well as clean-up adsorbent (1.5401 g), n-hexane:dichloromethane (11:89, v/v) as the eluting solvent (15.45 mL) and Na2SO4 (2.0 g) as dehydrating agent. Determination and quantification of OCPs and PCBs residues were carried out using a gas chromatograph equipped with an electron capture detector (GC-ECD). A mass spectrometric detector (GC-MS) in the selected ion monitoring (SIM) mode was also used for confirmation purposes. Method detection limits by GC-MS ranged from 0.42 to 0.87 ng g�1 and 0.51 to 1.35 ng g�1, for OCPs and PCBs, respectively. Lower detection limits were calculated for GC-ECD ranging between 0.15-0.30 ng g�1 and 0.18-0.48 ng g�1, respectively. Relative standard deviations did not exceed 9%. Analytes provided recoveries ranging from 65% to 106%. The proposed method was successfully applied to the analysis of lung tissues from six autopsy cases.

Περιγραφή

Λέξεις-κλειδιά

Matrix solid-phase dispersion, Organochlorine pesticides, Polychlorinated biphenyls, Post-mortem human lung, Experimental design, Desirability profile

Θεματική κατηγορία

Παραπομπή

Σύνδεσμος

http://www.sciencedirect.com/science/article/pii/S0021967312000076

Γλώσσα

en

Εκδίδον τμήμα/τομέας

Όνομα επιβλέποντος

Εξεταστική επιτροπή

Γενική Περιγραφή / Σχόλια

Ίδρυμα και Σχολή/Τμήμα του υποβάλλοντος

Πανεπιστήμιο Ιωαννίνων. Σχολή Θετικών Επιστημών. Τμήμα Χημείας

Πίνακας περιεχομένων

Χορηγός

Βιβλιογραφική αναφορά

Ονόματα συντελεστών

Αριθμός σελίδων

Λεπτομέρειες μαθήματος

item.page.endorsement

item.page.review

item.page.supplemented

item.page.referenced