Protrusio of a ceramic femoral head through the acetabular metallic shell, extensive metallosis and 'bubble sign'

Φόρτωση...
Μικρογραφία εικόνας

Ημερομηνία

Συγγραφείς

Malizos, K.
Roidis, N. T.
Poultsides, L.
Basdekis, G.
Moraitis, T.
Xenakis, T.

Τίτλος Εφημερίδας

Περιοδικό ISSN

Τίτλος τόμου

Εκδότης

Περίληψη

Τύπος

Είδος δημοσίευσης σε συνέδριο

Είδος περιοδικού

peer-reviewed

Είδος εκπαιδευτικού υλικού

Όνομα συνεδρίου

Όνομα περιοδικού

Orthopedics

Όνομα βιβλίου

Σειρά βιβλίου

Έκδοση βιβλίου

Συμπληρωματικός/δευτερεύων τίτλος

Περιγραφή

A 24-year-old patient with a history of juvenile rheumatoid arthritis underwent a primary cementless left total hip arthroplasty (THA). The original THA consisted of an Optifix 54 cup with a 3-mm thick polyethylene liner, an Optifix size 4 stem (Smith & Nephew Richards, Memphis, Tennessee) and a Biolox aluminum 32-mm femoral head. Fourteen years later, radiographs demonstrated extensive wear of the polyethylene liner resulting in direct articulation and abrasion wear of the ceramic femoral head on the cup and a bubble sign. This article presents a case of a catastrophic failure of a ceramic/polyethylene bearing with destruction of the polyethylene liner and the metallic shell and protrusio of the nonfractured ceramic head through the metallic shell. To our knowledge this is the first description of extensive metallosis and subsequent radiograph bubble sign not presenting as a result of wear of a metal-on-metal articulation. At the time of revision surgery-Hydrocel TNT Monoblock 58 cup (Zimmer, Warsaw, Indiana), Wagner 265/14 stem (Zimmer), and a Co/Cr 28-mm head-copious metallic debris was seen both macroscopically and histologically, with the ceramic head protruding behind the metallic shell. Multiple factors may have been responsible for this failure including a thin polyethylene shell, a suboptimal locking mechanism, gamma in air sterilization for polyethylene, multiple screw-holes that reduce the contact surface between shell and polyethylene, the rough surface on the inside of the shell and non-articular wear at the metal polyethylene interface within the acetabular component and the high demands of this active young patient.

Περιγραφή

Λέξεις-κλειδιά

Arthroplasty, Replacement, Hip/*adverse effects, Ceramics, Hip Joint/radiography, Humans, Male, Metals/adverse effects, *Prosthesis Failure, Reoperation, Young Adult

Θεματική κατηγορία

Παραπομπή

Σύνδεσμος

http://www.ncbi.nlm.nih.gov/pubmed/19301791

Γλώσσα

en

Εκδίδον τμήμα/τομέας

Όνομα επιβλέποντος

Εξεταστική επιτροπή

Γενική Περιγραφή / Σχόλια

Ίδρυμα και Σχολή/Τμήμα του υποβάλλοντος

Πανεπιστήμιο Ιωαννίνων. Σχολή Επιστημών Υγείας. Τμήμα Ιατρικής

Πίνακας περιεχομένων

Χορηγός

Βιβλιογραφική αναφορά

Ονόματα συντελεστών

Αριθμός σελίδων

Λεπτομέρειες μαθήματος

item.page.endorsement

item.page.review

item.page.supplemented

item.page.referenced