Construction and analytical applications of a palm-sized microcontroller-based amperometric analyzer

dc.contributor.authorAvdikos, E. M.en
dc.contributor.authorProdromidis, M. I.en
dc.contributor.authorEfstathiou, C. E.en
dc.date.accessioned2015-11-24T16:54:46Z
dc.date.available2015-11-24T16:54:46Z
dc.identifier.issn0925-4005-
dc.identifier.urihttps://olympias.lib.uoi.gr/jspui/handle/123456789/10196
dc.rightsDefault Licence-
dc.subjectamperometric analyzeren
dc.subjectpotentiostaten
dc.subjectmicrocontrollersen
dc.subjectascorbic acid assayen
dc.subjectblooden
dc.subjectpotentiostaten
dc.titleConstruction and analytical applications of a palm-sized microcontroller-based amperometric analyzeren
heal.abstractThe design and development of a palm-sized (9 cm x 11 cm x 3 cm), cost-effective, microcontroller-operated analyzer for direct amperometric measurements is described. The low-power-consumption electronics used allow 8 h of autonomous operation with a 9 V battery (110 mAh), making thus this unit suitable for in-field measurements. Its operation is based mainly on the simple two-electrode potentiostatic mode, although the three-electrode mode is an option. The use of a microcontroller combined with analog and digital supporting circuits allows: (i) generation of the applied potential and the acquisition of analog signals with a gain auto-scaling capability; (ii) interaction with the operator for setting the measurement parameters; (iii) processing of data and displaying result on an LCD screen. Numerical data are automatically stored in memory and they can be retrieved by a personal computer through an RS232 port, either for creating measurements archives or for advanced processing. For this purpose, a program has been developed (on the LabVIEW platform) to provide a user-friendly graphical interface. The utility of this amperometric analyzer was assessed by performing experiments for the determination of ascorbic acid in standard solutions and pharmaceutical tablets. (c) 2004 Elsevier B.V. All rights reserved.en
heal.accesscampus-
heal.fullTextAvailabilityTRUE-
heal.identifier.primaryDOI 10.1016/j.snb.2004.10.029-
heal.identifier.secondary<Go to ISI>://000229665700053-
heal.identifier.secondaryhttp://ac.els-cdn.com/S0925400504007221/1-s2.0-S0925400504007221-main.pdf?_tid=91ccc700167b2f6529f6acb9b9affa7e&acdnat=1333037128_2d96ed98729237792a93484f16836d7e-
heal.journalNameSensors and Actuators B-Chemicalen
heal.journalTypepeer reviewed-
heal.languageen-
heal.publicationDate2005-
heal.publisherElsevieren
heal.recordProviderΠανεπιστήμιο Ιωαννίνων. Σχολή Θετικών Επιστημών. Τμήμα Χημείαςel
heal.typejournalArticle-
heal.type.elΆρθρο Περιοδικούel
heal.type.enJournal articleen