Assessment of the performances of three oscillometric blood pressure monitors for neonates using a simulator
| dc.contributor.author | Papadopoulos, G. | en |
| dc.contributor.author | Mieke, S. | en |
| dc.contributor.author | Elisaf, M. S. | en |
| dc.date.accessioned | 2015-11-24T19:35:08Z | |
| dc.date.available | 2015-11-24T19:35:08Z | |
| dc.identifier.issn | 1359-5237 | - |
| dc.identifier.uri | https://olympias.lib.uoi.gr/jspui/handle/123456789/23681 | |
| dc.rights | Default Licence | - |
| dc.subject | Blood Pressure/physiology | en |
| dc.subject | *Blood Pressure Monitors | en |
| dc.subject | Humans | en |
| dc.subject | Infant, Newborn/*physiology | en |
| dc.title | Assessment of the performances of three oscillometric blood pressure monitors for neonates using a simulator | en |
| heal.abstract | BACKGROUND: The majority of monitors for non-invasive measurement of blood pressure in neonates and infants use the oscillometric method. OBJECTIVE: To use a simulator to investigate the overall system accuracy for three oscillometric blood pressure monitors. The tested devices are DINAMAP 8100 (Critikon), SpaceLabs M90426 (SpaceLabs Medical)and the module HP M 1008B (Hewlett-Packard). METHODS: The blood pressure values obtained by the three devices were compared with those of the invasive reference. A blood pressure simulator was used for testing the performance of the three blood pressure monitors. RESULTS: The Dinamap and SpaceLabs readings are generally in good agreement with the invasive reference. In contrast, the HP M 1008B readings for the diastolic and mean arterial pressures are inaccurate (the mean errors are 21 and 15mmHg, respectively). CONCLUSION: These results contradict our previously published data obtained when a different simulator (Cufflink)was used. These findings are important, because blood pressure values that are too low can result in a different therapeutic approach being applied, for newborn babies have only a small range of mean arterial pressure for the autoregulation of the cerebral blood flow. | en |
| heal.access | campus | - |
| heal.fullTextAvailability | TRUE | - |
| heal.identifier.secondary | http://www.ncbi.nlm.nih.gov/pubmed/10362888 | - |
| heal.journalName | Blood Press Monit | en |
| heal.journalType | peer-reviewed | - |
| heal.language | en | - |
| heal.publicationDate | 1999 | - |
| heal.recordProvider | Πανεπιστήμιο Ιωαννίνων. Σχολή Επιστημών Υγείας. Τμήμα Ιατρικής | el |
| heal.type | journalArticle | - |
| heal.type.el | Άρθρο Περιοδικού | el |
| heal.type.en | Journal article | en |
Αρχεία
Φάκελος/Πακέτο αδειών
1 - 1 of 1
Φόρτωση...
- Ονομα:
- license.txt
- Μέγεθος:
- 1.74 KB
- Μορφότυπο:
- Item-specific license agreed upon to submission
- Περιγραφή: