heal.abstractWe demonstrate by ray tracing analysis that the focusing properties of a hemispherical spectrograph depend critically on the placement of the electron entrance position R(0) and the value of the potential V(0) at R(0). Improvements over conventional spectrographs using R(0) = (R) over bar, the mean radius and V(0) = 0 are predicted for particular combinations of R(0) and V(0). A spectrograph with (R) over bar = 101.6 mm and an entrance at R(0) = 82.55 mm using a zoom lens and a two-dimensional position sensitive detector (2D-PSD) with 40 mm multichannel plates and resistive anode encoder has been used to confirm the predicted qualitative behaviour. These results suggest that improvement in energy resolution may be attained without the use of fringing field correctors. (C) 2000 Elsevier Science B.V. All rights reserved.en
heal.accesscampus-
heal.fullTextAvailabilityTRUE-
heal.identifier.primaryDoi 10.1016/S0168-9002(99)00954-7-
heal.identifier.secondary<Go to ISI>://000085077600021-
heal.identifier.secondaryhttp://ac.els-cdn.com/S0168900299009547/1-s2.0-S0168900299009547-main.pdf?_tid=8b3e4ce2-4de8-11e3-9a5d-00000aacb362&acdnat=1384514871_7bf18fc797e091a8a2cd7b657bb088a1-
heal.journalNameNuclear Instruments & Methods in Physics Research Section a-Accelerators Spectrometers Detectors and Associated Equipmenten
heal.journalTypepeer reviewed-
heal.languageen-
heal.publicationDate2000-
heal.recordProviderΠανεπιστήμιο Ιωαννίνων. Σχολή Επιστημών και Τεχνολογιών. Τμήμα Βιολογικών Εφαρμογών και Τεχνολογιώνel
heal.typejournalArticle-
heal.type.elΆρθρο Περιοδικούel
heal.type.enJournal articleen

Αρχεία

Φάκελος/Πακέτο αδειών

Προβολή: 1 - 1 of 1
Φόρτωση...
Μικρογραφία εικόνας
Ονομα:
license.txt
Μέγεθος:
1.74 KB
Μορφότυπο:
Item-specific license agreed upon to submission
Περιγραφή: