The effect of altering angulation in T loops on the moment to force ratio
dc.contributor.author | Nakos, G. | en |
dc.date.accessioned | 2015-11-24T19:42:04Z | |
dc.date.available | 2015-11-24T19:42:04Z | |
dc.identifier.issn | 0898-7394 | - |
dc.identifier.uri | https://olympias.lib.uoi.gr/jspui/handle/123456789/24581 | |
dc.rights | Default Licence | - |
dc.subject | Dental Stress Analysis | en |
dc.subject | *Orthodontic Appliances | en |
dc.subject | *Orthodontic Wires | en |
dc.title | The effect of altering angulation in T loops on the moment to force ratio | en |
heal.abstract | The effect of changing angulation of a T loop on the force systems developed at the alpha and beta ends was determined. Standard 0.017'' X 0.25'' tMA T loops were used. The force systems were determined using testing apparatus developed in this laboratory. The relationship between moment to force ratio and activation is linear at higher activations (3mm-7mm), and becomes non-linear at lower activations (less than 3mm). The relationship between moment to force ratio and spring angulation is also linear. The clinical application of these findings is discussed. | en |
heal.access | campus | - |
heal.fullTextAvailability | TRUE | - |
heal.identifier.secondary | http://www.ncbi.nlm.nih.gov/pubmed/3273264 | - |
heal.journalName | Conn Dent Stud J | en |
heal.journalType | peer-reviewed | - |
heal.language | en | - |
heal.publicationDate | 1988 | - |
heal.recordProvider | Πανεπιστήμιο Ιωαννίνων. Σχολή Επιστημών Υγείας. Τμήμα Ιατρικής | el |
heal.type | journalArticle | - |
heal.type.el | Άρθρο Περιοδικού | el |
heal.type.en | Journal article | en |